There are 9 good ways to handle faults in the magnetic flip plate level gauge to share with you: RUNDE speciallized in bronze fittings,covered end feed and solder type and the flange,Bronze Fittings.well known as yorkshiire bronze fittings,pegler bronze or gunmetal fittings as well as connex bronze fittings bronze fittings,Yorkshire bronze fittings,Pegler bronze fittings,gunmetal fittings Taizhou Runde Company , https://www.zjbronzefittings.com
1, state adjustment method
In general, do not touch the components in the circuit before the fault is determined. This is especially true for adjustable devices such as potentiometers. However, if multiple reference measures are taken in advance (for example, position marks or voltage values ​​or resistance values ​​are measured before being touched), they are allowed to act if necessary. Maybe after some change, the fault will be eliminated.
2, isolation method
Fault isolation method does not require the same type of equipment or spare parts for comparison, and it is safe and reliable. According to the fault detection flow chart, segmentation and enveloping gradually narrow down the fault search range, together with signal comparison, component exchange and other methods, the fault will generally be quickly found.
3, capacitor bypass method
When a circuit produces a strange phenomenon, such as a disorganized display, the capacitor bypass method can be used to determine the portion of the circuit that has failed. Connect the capacitor across the power and ground of the ic. Connect the transistor circuit across the base or collector output and observe the effect on the fault. If the fault phenomenon disappears when the capacitor bypass input is inactive and its output is bypassed, a fault is identified in this stage of the circuit.
4, observation method
Uses sight, smell, touch. In some cases, the damaged components will discolor, blistering, or burn spots; burned devices will produce some special odors; short-circuited chips will become hot; visible or unwelded areas can also be observed with the naked eye. .
5, elimination method
The so-called exclusion method is to determine the cause of the fault by inserting and removing some plug-in boards and devices. When the instrument returns to normal after removing a certain board or device, it indicates that the failure occurred.
6, knocking hand pressure method
The phenomenon of good or bad operation of the instrument is often encountered. The vast majority of this phenomenon is due to poor contact or welding. For this situation, tapping and hand pressure can be used.
The so-called "knocking" is to lightly hit the board or component with a small rubber hoe or other knocker to see if it will cause an error or stop failure. The so-called "hand pressure" is when the fault occurs, after the power is turned off, the plugged parts and the plug and the seat are re-pressed firmly by hand, and if the power supply is turned on, it will eliminate the fault. If it is found that when the cabinet is normal and the beat is not normal, all joints should be re-inserted and tried again. If the brains are not successful, you have to find another solution.
7, lift temperature method
Sometimes, the instrument work for a long time, or in the summer when the working environment temperature is high, it will malfunction, shut down and check normally, stop for a period of time and then boot again and normal, after a while there is a failure. This phenomenon is due to the poor performance of individual ICs or components, and the high temperature characteristic parameters do not meet the requirements of the specifications. In order to find out the cause of the fault, it is possible to use the heating and cooling method.
The so-called cooling, that is, in the event of a failure, the use of cotton fiber will be wiped in the area of ​​possible failure of the alcohol to make it cool, observe whether the failure is eliminated. The so-called warming is artificially raising the temperature of the environment, such as using a soldering iron near the point of doubt (note that the temperature must not rise too high so as to damage the normal device) to see if the fault appears.
8, replacement method
Two identical instruments or enough spare parts are required. Replace a good spare with the same component on the faulty machine and see if the fault is eliminated.
9, riding a law
The riding method is also called parallel method. Place a good IC chip on the chip to be inspected, or connect good components (resistance capacitors, diodes, transistors, etc.) in parallel with the components to be inspected and maintain good contact if the fault is caused by an open circuit inside the device or For poor contact and other reasons, this method can be ruled out.
10, magnetic flap level gauge comparison method
Two instruments of the same type are required and one is in normal operation. Use this method also has the necessary equipment, for example, multimeter, oscilloscope and so on. According to the nature of comparison, there are voltage comparison, waveform comparison, static impedance comparison, output comparison, and current comparison. The specific method is: let the faulty meter and the normal meter operate under the same conditions, and then detect the signals of some points and then compare the measured two signals. If they are different, you can conclude that the fault is out here. This method requires maintenance personnel with considerable knowledge and skills.